The sample manager uses a direct injection mechanism to inject samples drawn from vials and plates onto a chromatographic column..
The Alliance iS Sample Manager maintains sample temperature, aspirates precise samples, and then injects them into the mobile phase using a flow-through-needle (FTN) design. The needle becomes part of the fluidics path when the sample is injected onto the column. The mobile phase cleanses the needle during the run, ensuring complete sample recovery and minimizing carryover. Additionally, the sample positioning mechanism controls the location of the sample plates relative to the sample needle, reducing extra system volume.
The sample manager features:
Need Help? Connect with a Waters expert.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.
Refer to the English version of this page for the latest information.