Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs! Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs!
A leak sensor in the drip tray continuously monitors the detector for leaks.
Warning: To avoid personal contamination with biologically hazardous or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
The detector's leak sensor stops system flow when it detects accumulated, leaked liquid in its reservoir. When the sensor detects a leak, the system's touchscreen displays an alarm message.
Required tools and materials
Chemical-resistant, powder-free gloves
Replacement leak sensor
To replace the detector's leak sensor:
Open the detector's door, gently pulling its right-hand edge toward you.
Remove the old leak sensor:
To detach the leak sensor connector from the front of the instrument, press down on the release tab.
Note: The following figure is for representation only. Your hardware may differ slightly from that shown.
Release tab
Leak sensor connector
To remove the leak sensor from the reservoir, grasp it by the serrations around its top and pull upward on it (see Figure "Installed leak sensor, rear view").
Install the new leak sensor:
Holding the leak sensor by its serrations, align its T-bar with the slot in the side of the reservoir and slide it into place (see Figure "Installed leak sensor, rear view").
Serrations
T-bar engaged with reservoir slot
Leak sensor
Plug the leak sensor connector into the front of the instrument.
Close the detector's door.
From the system's touchscreen Commands view, tap Reset.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.
Refer to the English version of this page for the latest information.