Observing vial and plate recommendations

Observing vial and plate recommendations
Topic LCI-USG-0001, Last Updated 11/08/2024

Incorrect selection of sample vials and plates can lead to issues with system function and performance.

Waters recommends that you observe these usage guidelines for sample vials and plates in the sample manager:

  • Vials
    • Use only Waters-certified vials.
    • Ensure that vial holders conform to ANSI/SBS standards.
  • Plates
    • Use only Waters-approved plates and cap mats.
    • When selecting a new plate supplier, especially for 384-well plates, measure the plate size to ensure compatibility with Waters’ specifications for the sample manager.
    • To avoid warping plates, do not centrifuge them.
    • Be aware that plates containing samples with high concentrations of organic solvent can give inconsistent results at or above room temperature due to solvent evaporation.
  • Covers
    • Use cap mats on sample plates whenever possible.
    • Use pre-slit cap mats/seals and vial caps. Use of non-pre-slit cap mats and vial caps can cause clogging in the wash drainage lines.
    • To prevent sample spillage or needle damage, use only Waters-approved covers on the sample vials.
See also: Waters Sample Vials and Accessories Brochure (720001818EN), or visit https://preprod1.waters.com/nextgen/us/en/products/vials--plates--and-certified-containers.html?icid=hm-fea_00512 for information on plates and vials.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.

Back To Top Back To Top