Metallic fitting with short or long flats and two-piece metallic ferrule (V-detail)

Metallic fitting with short or long flats and two-piece metallic ferrule (V-detail)
Topic LCI-USG-0106, Last Updated 11/08/2024

The procedure for tightening metallic fittings differs for new versus reused fittings.

In order to avoid leaks, see the following recommendations in this guide:

Note: Before tightening compression screws, ensure that the tubing is fully bottomed in its connection port.
Note: Different techniques apply to installing new fittings and re-tightening previously used fittings.

First use

Note: Tighten the fitting finger-tight plus an additional 3/4-turn using a 1/4-inch open-end wrench.
Figure 1: Metallic fitting with short flats and two-piece metallic ferrule, first use
Short flats
Compression screw
Two-piece metallic ferrule
Figure 2: Metallic fitting with long flats and two-piece metallic ferrule, first use
Long flats
Compression screw
Two-piece ferrule
Figure 3: Metallic fitting with short or long flats and two-piece metallic ferrule, first-use tightening

Reinstallation

Note: For best performance, only reinstall this connection into the same port from which it was removed.

Tighten the fitting finger-tight plus as much as an additional 1/6-turn using a 1/4-inch open-end wrench.

Figure 4: Metallic fitting with short flats and two-piece metallic ferrule, reinstallation
Short flats
Compression screw
Two-piece metallic ferrule
Figure 5: Metallic fitting with long flats and two-piece metallic ferrule, reinstallation
Long flats
Compression screw
Two-piece ferrule
Figure 6: Metallic fitting with short or long flats and two-piece metallic ferrule, reinstallation tightening

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Refer to the English version of this page for the latest information.

Back To Top Back To Top