Testing the PDA detector for noise and drift

Testing the PDA detector for noise and drift
Topic LCI-USG-0220, Last Updated 11/07/2024

A dirty flow cell or a flow cell with bubbles can cause excessive noise and drift.

You can run a noise and drift test on the PDA detector from the system touchscreen.

  1. From the touchscreen, tap HEALTH > Troubleshoot.
  2. Tap the detector icon.
  3. Tap Noise and Drift Test, follow the on-screen prompts, and then tap START to begin the test.
    The screen reads Noise and Drift Test Completed and displays the test results.
  4. Tap DONE.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.