Single-Wavelength mode

Single-Wavelength mode
Topic LCI-USG-0084, 最近更新 11/22/2024

Single-Wavelength mode is the detector's default mode of operation.

The detector supports monitoring of a single wavelength from 190 nm to 700 nm, settable in 1-nm increments on channel A.

In this mode, the detector automatically engages the optical, second-order filter for wavelengths 370 nm and above and removes it for wavelengths under 370 nm. This filter blocks unwanted UV light from striking the diffraction grating and interfering with absorbance detection above 370 nm.

When using the detector in this mode, you can configure the additional parameters listed in the following table and set up an Events table for the detector.

Table 1. Single-Wavelength mode configurable parameters
Parameter Description
Lamp Set the detector's lamp to On or Off
Data rate (Hz) Specify a sampling frequency up to 160 Hz (single channel, 2 Hz for dual channel)
Wavelength A (nm) Specify a wavelength for channel A, from 190 nm to 700 nm, settable in 1-nm increments
Note: If the detector is in Dual-Wavelength mode, the value of Wavelength B also appears.
Filter time constant (s) Program a filter time. Options are No Filter, Slow, Normal, Fast, and Custom (see Filtering noise for detailed information about this parameter, these setting options, and the different effects produced).
Autozero (two options):
  • Autozero on Inject Start
  • Autozero on Wavelength Change
Set when to autozero (resets the detector offsets)

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首