Preventing leaks

Preventing leaks
Topic LCI-USG-0099, 最近更新 11/22/2024

Preventing leaks during an analysis ensures adequate flow pressure in the system and the integrity of the sample.

Leaks can occur at any tubing connection, gasket, or seal, but they are most common at tubing connections. Low-pressure leaks (on the intake side of the solvent manager’s pump) cause solvent loss and air introduction during the intake cycle. Leaks at high-pressure fittings (downstream of the check valves) can release solvent but do not introduce air.

To prevent leaks, follow Waters’ recommendations for the proper tightening of system fittings. Note that different techniques apply to re-tightening fittings and installing them for the first time.


需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首