The Empower CDS provides features that are specific to the Alliance iS HPLC System. For more information, see the following topics, Intended use of the system, and the Empower online Information System.
需要協助嗎?聯絡Waters專家。
為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。