Replacing the pump mixer

Replacing the pump mixer
Topic LCI-USG-0189, 最近更新 11/22/2024

Users or Waters field service personnel can replace the pump mixer.

Warning: Observe Good Laboratory Practice (GLP) at all times, particularly when working with hazardous materials. Consult the Safety Data Sheets regarding the solvents you use. Additionally, consult the safety representative for your organization regarding its protocols for handling such materials.
Note: To avoid contaminating system components, wear clean, chemical-resistant, powder-free gloves when performing this procedure.

Required tools and materials

  • Chemical-resistant, powder-free gloves
  • 1/4-inch open-end wrench
  • 3/8-inch open-end wrench
  • Replacement mixer
Note: There is an optional 690-μL diffusion bonded mixer kit that offers significantly better mixing performance than the traditional 675-μL or 680-μL mixers, resulting in lower compositional noise and a quieter baseline in select applications. For information on installing the 690-μL diffusion bonded mixer, see the Ti Diffusion Bonded Mixer - 690 μL Kit Installation Guide (715009251).

To replace the mixer:

  1. Flush the pump with nonhazardous solvent.
  2. Stop the solvent flow.
  3. Open the pump compartment door.
  4. Remove the mixer from the mixer clip.
  5. Using the 3/8-inch open-end wrench to hold the mixer in place, disconnect the outlet compression fitting using the 1/4-inch open-end wrench.
    Figure 1: Location of the outlet compression fitting for the mixer
    Mixer
    Wrench flats
    Outlet compression fitting
  6. Using the 3/8-inch open-end wrench to hold the mixer, disconnect the inlet compression fitting using the 1/4-inch wrench.
    Figure 2: Location of the inlet compression fitting for the mixer
    Inlet compression fitting
    Wrench flats
    Mixer
  7. Unpack the replacement mixer.
    Note: Ensure that the arrow on the mixer is pointing from left to right.
  8. Reattach the compression fittings to the mixer and tighten them finger-tight, plus as much as an additional 1/6-turn for existing fittings or 1/2-turn for new fittings.
  9. Insert the body of the mixer into the clips.
  10. Close the pump compartment door.
  11. From the touchscreen Commands view, tap Reset.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首