Observing vial and plate recommendations

Observing vial and plate recommendations
Topic LCI-USG-0001, 最近更新 11/22/2024

Incorrect selection of sample vials and plates can lead to issues with system function and performance.

Waters recommends that you observe these usage guidelines for sample vials and plates in the sample manager:

  • Vials
    • Use only Waters-certified vials.
    • Ensure that vial holders conform to ANSI/SBS standards.
  • Plates
    • Use only Waters-approved plates and cap mats.
    • When selecting a new plate supplier, especially for 384-well plates, measure the plate size to ensure compatibility with Waters’ specifications for the sample manager.
    • To avoid warping plates, do not centrifuge them.
    • Be aware that plates containing samples with high concentrations of organic solvent can give inconsistent results at or above room temperature due to solvent evaporation.
  • Covers
    • Use cap mats on sample plates whenever possible.
    • Use pre-slit cap mats/seals and vial caps. Use of non-pre-slit cap mats and vial caps can cause clogging in the wash drainage lines.
    • To prevent sample spillage or needle damage, use only Waters-approved covers on the sample vials.
See also: Waters Sample Vials and Accessories Brochure (720001818EN), or visit https://preprod1.waters.com/nextgen/us/en/products/vials--plates--and-certified-containers.html?icid=hm-fea_00512 for information on plates and vials.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首