Shutting down for more than 24 hours

Shutting down for more than 24 hours
Topic LCI-USG-0214, 最近更新 11/22/2024

Failure to flush the system/detector before shutdown can result in a plugged flow cell.

To shut down the detector for more than 24 hours:

  1. Extinguish the detector lamp by tapping COMMANDS > UV Detector Lamp.
    Power off Lamp appears on the screen.
  2. Remove buffer salts and additives by flushing with water.
  3. Flush the column and flow cell with 100% pure organic solvent.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首