Verify the detector calibration after removing and replacing a flow cell or if verification fails on system startup.
To verify the PDA detector's calibration, prime the system and run flow for ten minutes to flush the flow cell with solvent, and ensure that it is fully wetted. A 90:10 water/acetonitrile mixture is recommended at 0.5 mL/min. Warm up the deuterium lamp for a minimum of five minutes, and ensure that it is in the "ON" state prior to verification.
If buffers were recently used in the flow cell, Waters recommends that you first flush with 10 mL of HPLC-grade water, followed by 10 mL of a low-surface-tension solvent such as methanol (provided it is miscible with the previous mobile phase).
To verify detector calibration:
需要協助嗎?聯絡Waters專家。
為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。