Mercury calibration

Mercury calibration
Topic LCI-USG-0207, 最近更新 11/22/2024

Mercury calibration uses a mercury-argon spectral calibration lamp to calibrate or recalibrate the instrument. Mercury calibration is typically performed at the factory but can be performed by qualified Waters field service engineers if necessary. Note that mercury recalibration should only be performed if erbium calibration fails or the prior mercury calibration parameters were wiped or corrupted.

A Waters field service engineer can perform a mercury calibration for the PDA detector from the system touchscreen.

Important:
  • Mercury calibration is to be performed only by a Waters field service engineer.
  • Impurities in the flow cell can affect wavelength calibration. Ensure that the flow cell is clean before you perform calibration.
  • This procedure can adversely affect spectral library matching and peak purity calculations.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首